FDA Adverse Event Malfunction Summary report: N

COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER

MDR report key: 3982397 · Received August 6, 2014

Report

Report Number
1061932-2014-01824
Event Type
Malfunction
Date Received
August 6, 2014
Date of Event
July 13, 2014
Report Date
July 13, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K010765
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT. THE FSE OBSERVED A HOLE IN THE TUBING AT PINCH VALVE (PV45). THE FSE REPLACED THE TUBING AT PINCH VALVE (PV45), RESOLVING THE LEAK ISSUE. THE FSE VERIFIED THE INSTRUMENT AND THE INSTRUMENT WAS RETURNED TO NORMAL OPERATION. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED LESS THAN 1 ML OF DILUENT MIXTURE LEAKED FROM THE COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER. THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT. THERE WAS NO EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES. THE CUSTOMER WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF GLOVES, GOGGLES AND A LAB COAT AT THE TIME OF THE OCCURRENCE. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION WITH THE REPORTED EVENT. THERE WAS NO DEATH, INJURY, OF AFFECT TO USER OR TO PATIENT TREATMENT ATTRIBUTED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
461887 COULTER HMX HEMATOLOGY ANALYZER WITH AUTOLOADER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1