FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3982355 · Received August 5, 2014

Report

Report Number
3004209178-2014-90099
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 3, 2014
Report Date
July 5, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH MOTOR ERROR ALARM DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. UNABLE TO PERFORM NO DELIVERY TEST OR DISPLACEMENT TEST DUE TO MOTOR ERROR ALARM. THE INSULIN PUMP HAD MINOR SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, CRACKED BELT CLIP SLOT AT THE BATTERY TUBE THREADS AREA AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER REPORTED ISSUES WITH THE INSULIN PUMP. THE BLOOD GLUCOSE READING IS 102 MG/DL. CUSTOMER STATES THAT THE INSULIN DIDN'T EXIT AND THE PUMP ALARMED NO DELIVERY. CUSTOMER WENT THROUGH THE DISPLACEMENT TEST PROCESS AND THE TEST FAILED. THE INSULIN PUMP DID NOT ALARM LOW RESERVOIR. NOTHING FURTHER REPORTED

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
458384 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 27 YR