FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3982348 · Received August 5, 2014

Report

Report Number
3004209178-2014-90255
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 5, 2014
Report Date
July 5, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP HAD MOISTURE DAMAGE WAS FOUND ON KEYPAD TRACES. ALL BUTTONS FUNCTION PROPERLY AND NO FROZEN SCREEN NOTED. INSULIN PUMP PASSED UNEXPECTED AND SELF TEST. NO ALARMS NOTED. INSULIN PUMP WAS MONITORED FOR 24 HOURS WITH MULTIPLE BASAL RATES. NO ALARMS DURING TESTING. AN ALARM WAS IN PUMP HISTORY DUE TO CORRUPTED HISTORY FILE. INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE SCREEN OF THE INSULIN PUMP BECAME FROZEN. THE BLOOD GLUCOSE READING WAS 265 MG/DL. THE INSULIN PUMP GAVE OFF A SAFETY ALARM DUE TO BEING STORED OR NOT IN USE FOR AN EXTENDED PERIOD OF TIME. CUSTOMER ALSO REPORTED THAT THE BUTTONS BECAME UNRESPONSIVE. SHE COULD NOT RECALL ANY SIGNIFICANT EVENTS LEADING TO THIS KEYPAD ISSUE. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
457667 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAP

Patients

Seq Age Sex Outcome Treatment
1 54 YR