FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3982311 · Received August 5, 2014

Report

Report Number
2032227-2014-06186
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
May 21, 2014
Report Date
May 21, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

UNIT RECEIVED INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO RAMPING NUMBERS ANOMALY NOTED. PUMP RECEIVED WITH SCRATCHED LCD WINDOW. UNIT RECEIVED WITH CRACKED CASE DISPLAY WINDOW CORNER. UNIT RECEIVED WITH CRACKED BATTERY TUBE THREADS. PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP. UNIT RECEIVED WITH CRACKED BELT CLIP SLOT.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE NUMBERS ON THE DISPLAY SCREEN OF THE INSULIN PUMP WERE RAMPING ON THEIR OWN. BLOOD GLUCOSE LEVEL WAS 162 MG/DL AT THE TIME OF THE CALL. NO FURTHER INFORMATION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
456585 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAS A000207835

Patients

Seq Age Sex Outcome Treatment
1 72 YR