FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3982275 · Received August 5, 2014

Report

Report Number
3004209178-2014-90264
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 5, 2014
Report Date
July 5, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING, IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED THE INSULIN PUMP ALARMED MOTOR ERROR THROUGH OUT THE DAY DURING BOLUS. CUSTOMER'S STATED THAT HIS GLUCOSE METER STATED HE WAS HIGH. CUSTOMER DOES NOT RECALL ANY SIGNIFICANT EVENT THAT MAY HAVE CAUSED THE DEVICE TO ALARM. THE DEVICE WAS NOT EXPOSED TO MRI. CUSTOMER WAS ADVISED TO DISCONNECT FROM THE DEVICE. CUSTOMER DOES NOT USE THE SENSOR FEATURE. CUSTOMER WAS ABLE TO REWIND THE DEVICE. CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
457001 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAL

Patients

Seq Age Sex Outcome Treatment
1 23 YR