PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3982231 · Received August 5, 2014

Report

Report Number
3004209178-2014-90152
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 5, 2014
Report Date
July 5, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP ALARMED PRIME DURING PRIME ALARM TEST DUE TO FAULTY FORCE SENSOR. INSULIN PUMP HAD A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON DISPLAY WINDOW AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED AN ERROR WHILE PERFORMING RESERVOIR CHANGE. THE INSULIN PUMP KEEPS GOING BACK TO THE PRIME PROCESS. THE CUSTOMER'S BLOOD GLUCOSE WAS 220 MG/DL. THE DRIVE SUPPORT CAP WAS RECESSED. THE DISCONTINUATION OF THE INSULIN PUMP WAS ADVISED. NOTHING FURTHER WAS REPORTED.