FDA Adverse Event
Malfunction
Summary report: N
COULTER LH 780 HEMATOLOGY ANALYZER
MDR report key: 3982211
·
Received August 5, 2014
Report
- Report Number
- 1061932-2014-01849
- Event Type
- Malfunction
- Date Received
- August 5, 2014
- Date of Event
- July 15, 2014
- Report Date
- July 15, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K061616
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ON (B)(4) 2014, A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) EVALUATED THE ANALYZER AND REPLACED THE PILOT ACTUATORS AND TUBING THROUGH VL33 AND VL67, RESOLVING THE VLS ERRORS. (B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTED AN ERROR CONDITION OF "VLS ERRORS" WAS GENERATED WHEN USING THE COULTER LH 780 HEMATOLOGY ANALYZER. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER'S FACILITY TO EVALUATE THE ANALYZER. THERE WAS NO REPORT OF ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 456682 | COULTER LH 780 HEMATOLOGY ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |