FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3982090 · Received August 5, 2014

Report

Report Number
2032227-2014-06099
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 5, 2014
Report Date
July 5, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP PASSED THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND NO DELIVERY TESTS. THE DEVICE WAS PROGRAMMED WITH MULTIPLE BOLUSES AND MONITORED. THE BOLUSES WERE DELIVERED AND RECORDED PROPERLY IN THE BOLUS HISTORY SCREEN. NO BOLUS HISTORY ANOMALY NOTED. THE UNIT HAD CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

THE CUSTOMER STATED THAT THEIR INSULIN PUMP STOPPED DELIVERING INSULIN FOR THE PAST COUPLE OF WEEKS. THE CUSTOMER'S BLOOD GLUCOSE WAS 400 MG/DL AND WAS NOT TREATED. THE CUSTOMER STATED THAT THEY HAVE A BACK-UP PLAN. THE CUSTOMER ALSO STATED THAT THE INSULIN PUMP WAS NOT RECORDING THE BOLUS. WHEN CHECKING THE HISTORY OF THE DEVICE, IT WAS NOTED THAT THE LAST TWO BOLUSES WERE NOT RECORDED. WHILE TROUBLESHOOTING FOR THE HIGH BLOOD GLUCOSE, IT WAS NOTED THAT THE DATE AND THE TIME WERE PROGRAMMED INCORRECTLY IN THE INSULIN PUMP. ASSISTANCE WAS PROVIDED TO CORRECT THE ISSUE. ALSO, FOUND NO DELIVERY ALARM IN THE ALARM HISTORY OF THE INSULIN PUMP. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
456577 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAS

Patients

Seq Age Sex Outcome Treatment
1 77 YR