FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3982051 · Received August 5, 2014

Report

Report Number
3004209178-2014-89809
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 4, 2014
Report Date
July 4, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH NO ACT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACE. THE INSULIN PUMP RECEIVED WITH MINOR SCRATCHES ON DISPLAY WINDOW, CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS AND MISSING END CAP STICKER. NO MOISTURE DAMAGE INSIDE INSULIN PUMP NOTED.

Description of Event or Problem · 1

THE CUSTOMER'S MOTHER REPORTED THAT A BUTTON WAS NOT WORKING ON THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE WAS 217 MG/DL AT THE TIME THE ISSUE WAS REPORTED. THE CUSTOMER' S MOTHER STATED THAT THE INSULIN PUMP MIGHT HAVE BEEN EXPOSED TO MOISTURE. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
456567 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAS A6523NASJ

Patients

Seq Age Sex Outcome Treatment
1 13 YR