FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3982049 · Received August 5, 2014

Report

Report Number
3004209178-2014-89807
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 4, 2014
Report Date
July 4, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT FAMILY MEMBER OR FRIEND

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP WAS RECEIVED WITH UNRESPONSIVE BUTTONS AND BUTTON ERROR ALARM DUE TO CORRODED KEYPAD TRACES. INSULIN PUMP HAD A CRACKED DISPLAY WINDOW CORNER AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER'S FRIEND REPORTED THAT THE CUSTOMER RECEIVED A BUTTON ERROR ALARM ON THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE WAS 150 MG/DL AT THE TIME THE ISSUE WAS REPORTED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
457307 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS B1723NASJ

Patients

Seq Age Sex Outcome Treatment
1 36 YR