FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3982042
·
Received August 5, 2014
Report
- Report Number
- 3004209178-2014-89110
- Event Type
- Malfunction
- Date Received
- August 5, 2014
- Date of Event
- July 2, 2014
- Report Date
- July 2, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
MOTOR ERROR ALARM DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. UNABLE TO PERFORM THE DISPLACEMENT TEST DUE TO MOTOR ERROR ALARM. INSULIN PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHES ON LCD WINDOW, CRACKED RESERVOIR TUBE, CRACKED RESERVOIR TUBE WINDOW, MISSING END CAP STICKER AND CRACKED BATTERY TUBE THREADS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A MOTOR ERROR ALARM RANDOMLY ON HER INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS AT 162 MG/DL. THE CUSTOMER ALSO STATED THAT SHE HAD A MRI DONE AN HOUR PRIOR TO THE MOTOR ERROR ALARMED BUT STATED THAT SHE HAD REMOVED THE PUMP AND PLACED IN ANOTHER ROOM BEFORE TAKING THE MRI. THE INSULIN PUMP WAS REPLACED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 456564 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-522NAP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR |