FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3982040 · Received August 5, 2014

Report

Report Number
3004209178-2014-89107
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. INSULIN PUMP RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS, RESERVOIR TUBE AND MINOR SCRATCHED DISPLAY WINDOW.

Description of Event or Problem · 1

THE CUSTOMER REPORTED ISSUES WITH THE ACT AND ESC KEYPAD BUTTONS ON THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS AT 180 MG/DL. THE CUSTOMER STATED THERE WERE NO SIGNIFICANT EVENTS THAT WOULD HAVE LED TO THE KEY ISSUES ON HER INSULIN PUMP. THE INSULIN PUMP WAS REPLACED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
456986 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAP

Patients

Seq Age Sex Outcome Treatment
1 40 YR