FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3982011 · Received August 5, 2014

Report

Report Number
3004209178-2014-90005
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 7, 2014
Report Date
July 7, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO RESERVOIR ALARM WAS NOTED DURING PERFORMANCE TESTING. CRACKED CASE AT THE DISPLAY WINDOW CORNER, CRACKED RESERVOIR TUBE LIP, SCRATCHED LCD WINDOW, AND CRACKED BATTERY TUBE THREADS WERE ALSO NOTED.THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP HAD AN UNRESPONSIVE KEYPAD. THE BLOOD GLUCOSE READING WAS 4.9 MMOL/L. THE CUSTOMER STATED THAT SHE WAS TRYING TO CHANGE THE INFUSION SET AND A NO RESERVOIR ALARM OCCURRED. THE CUSTOMER COULD NOT CLEAR THE ALARM. ADVISED DISCONTINUATION OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
458687 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-554WWS

Patients

Seq Age Sex Outcome Treatment
1