FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3982003 · Received August 5, 2014

Report

Report Number
3004209178-2014-90001
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 7, 2014
Report Date
July 7, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO MOISTURE DAMAGE ON THE KEYPAD TRACE. NO BUTTON ERROR ALARM WAS NOTED. THE INSULIN PUMP ALSO HAD A SCRATCHED LCD WINDOW, CRACKED DISPLAY WINDOW CORNERS AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE BLOOD GLUCOSE READING WAS UNAVAILABLE. THE CUSTOMER STATED THAT THE ALARM OCCURRED WHEN ASLEEP. ADVISED CUSTOMER TO REMOVE THE BATTERY FOR 2 HOURS AND REST THE DEVICE. ADVISED REPLACEMENT OF THE INSULIN PUMP. NOTHING FURTHER REPORTED.