FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3981993 · Received August 5, 2014

Report

Report Number
2032227-2014-05428
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THERE WAS NO BUTTON ERROR ALARM NOTED. THE UNIT HAD NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THERE WAS NO UNEXPECTED NUMBERS RAMPING NOTED. THE UNIT HAD MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE AT DISPLAY WINDOW CORNER, SCRATCHED RESERVOIR TUBE WINDOW AND CRACKED RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER'S MOTHER CALLED TO REPORT A BUTTON ERROR ALARM DURING BOLUS. SHE STATES THAT THE BATTERIES WERE REPLACED BUT STILL HAD THE SAME ISSUE. THE INSULIN PUMP HAS BEEN REPLACED. NO ADDITIONAL INFORMATION IS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
457631 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-754LWWB

Patients

Seq Age Sex Outcome Treatment
1