FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3981970 · Received August 5, 2014

Report

Report Number
2032227-2014-05438
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 2, 2014
Report Date
July 2, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP ALARMED MOTOR ERROR DURING REWIND DUE TO MOTOR ENCODER SIGNAL OUT OF PHASE. INSULIN PUMP WAS UNABLE TO PERFORM DISPLACEMENT TEST DUE TO MOTOR ERROR ALARM. INSULIN PUMP HAD MINOR SCRATCHES ON DISPLAY WINDOW AND CRACKED DISPLAY WINDOW CORNER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A MOTOR ERROR ALARM ON THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE WAS 261 MG/DL. THE CUSTOMER STATED THAT THERE WAS NO SIGNIFICANT EVENT THAT COULD HAVE LED TO THE MOTOR ERROR ALARM. THE CUSTOMERS INSULIN PUMP WAS REPLACED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
458049 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723RNAP