FDA Adverse Event Malfunction Summary report: N

COULTER LH 780 HEMATOLOGY ANALYZER

MDR report key: 3981947 · Received August 5, 2014

Report

Report Number
1061932-2014-01813
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 22, 2014
Report Date
July 22, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061616
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE DETERMINED THE LIQUID FOUND WAS DUE TO A SPILL FROM THE SECONDARY PROBE WASH; THE AREA WAS CLEANED AND THERE WERE NO FURTHER LEAKS. TO ADDRESS THE DIFFERENTIAL SAMPLE PRESSURE ERRORS, THE FSE REPLACED RED TUBING TO THE TOP OF DIFFERENTIAL MIX CHAMBER AND DIFFERENTIAL SAMPLE PRESSURE REGULATOR. ALIGNMENT WAS PERFORMED. THE CUSTOMER LATER REPORTED INTERMITTENT DIFFERENTIAL SAMPLE PRESSURE ERRORS AGAIN. THE FSE RETURNED AND REPLACED ACTUATORS ON VL89, 41, 45, 51, AND 54 AS WELL AS THE PERFORMANCE VERIFICATION TEST PRINTED CIRCUIT BOARD. THERE WERE NO FURTHER REPORTS OF ERRORS. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED A SMALL POOL OF LIQUID WHILE TROUBLESHOOTING A COULTER LH 780 HEMATOLOGY ANALYZER FOR INTERMITTENT DIFFERENTIAL SAMPLE PRESSURE ERRORS. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE INSTRUMENT OPERATOR WAS WEARING GLOVES, GOGGLES, AND A LAB COAT WHEN THE LEAK WAS FOUND. THERE WERE NO REPORTS OF BIOHAZARD EXPOSURE TO THE LEAK. PATIENT RESULTS WERE NOT AFFECTED AND THERE WAS NO IMPACT TO PATIENT TREATMENT IN CONNECTION WITH THE LEAK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
457577 COULTER LH 780 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1