AU640 CLINICAL CHEMISTRY ANALYZER
Report
- Report Number
- 9612296-2014-00115
- Event Type
- Malfunction
- Date Received
- August 5, 2014
- Date of Event
- July 11, 2014
- Report Date
- July 17, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- JJE
- PMA / PMN Number
- K961274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A BECKMAN COULTER (BEC) FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT. THE FSE REPLACED THE POTASSIUM AND CHLORIDE ELECTRODES AND ALL ISE TUBING. IN ADDITION, THE FSE RESEATED THE ELECTRODE STACK IN THE FLOW CELL, VERIFIED O-RINGS WERE PLACED PROPERLY, REMOVED BUBBLES FROM STACK AND INCOMING REFERENCE SOLUTION, TIGHTENED STACK MOUNTING BRACKET, CHECKED THE GROUNDING OF THE ISE SYSTEM, AND CLEANED THE ISE SYSTEM. UPON FOLLOW UP, THE CUSTOMER DID NOT REPORT ANY FURTHER ISSUES; THE CUSTOMER STATED THE ISE SYSTEM WAS NO LONGER USED TO TEST PATIENT SAMPLES.
THE CUSTOMER REPORTED AN AU640 CLINICAL CHEMISTRY ANALYZER GENERATED ERRONEOUS POTASSIUM (K) RESULTS FOR TWO PATIENT SAMPLES. THE SAMPLES WERE RETESTED ON THE SAME INSTRUMENT AND ON AN ALTERNATE AU680 CLINICAL CHEMISTRY ANALYZER IN THE SAME LABORATORY; RETEST RESULTS ON THE ALTERNATE INSTRUMENT ARE CONSIDERED CORRECT. THERE WERE NO ERRONEOUS RESULTS REPORTED OUT OF THE LABORATORY AND THERE WAS NO IMPACT TO PATIENT TREATMENT IN CONNECTION WITH THIS EVENT. THE CUSTOMER STATED ION SELECTIVE ELECTRODE (ISE) QUALITY CONTROL (QC) RESULTS WERE WITHIN FACILITY ESTABLISHED RANGES. THE CUSTOMER TROUBLESHOOTED THE INSTRUMENT BY REPLACING THE K ELECTRODE, WHICH RESULTED IN A DROP IN CALIBRATION SLOPES, INDICATIVE OF ELECTRODE ISSUES. THE CUSTOMER THEN DETACHED AND CLEANED THE SAMPLE D-POT AND FLOW CELL. AFTER REINSTALLING, THE ION SELECTIVE ELECTRODE (ISE) GENERATED THERMISTOR TEMPERATURE ERRORS. THE ISE WASTE DRAIN WAS CHECKED FOR BUILDUP AND NO ISSUES WERE FOUND, BUT THE CUSTOMER NOTED THE DRAIN WAS MAKING A STRANGE NOISE THAT WAS NOT TYPICAL. THE CUSTOMER TURNED OFF THE ISE SYSTEM AND REQUESTED SERVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 457845 | AU640 CLINICAL CHEMISTRY ANALYZER | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER | AU640WI-U8 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |