FDA Adverse Event Malfunction Summary report: N

COAGUCHEK ® XS SYSTEM

MDR report key: 3981784 · Received August 5, 2014

Report

Report Number
1823260-2014-05906
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
June 30, 2014
Report Date
September 3, 2014
Manufacturer
ROCHE DIAGNOSTICS
Product Code
GJS
PMA / PMN Number
K062925
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA.

Description of Event or Problem · 1

CALLER TESTED 2.7 INR ON THE COAGUCHEK XS SYSTEM WHILE A COMPARISON LAB RETURNED AS 1.9 INR. THE CALLER'S COUMADIN DOSE WAS INCREASED BASED ON THE LAB VALUE. NO ADVERSE EVENT REPORTED. REQUESTED RETURN OF SUSPECT SYSTEM AND REPLACEMENT WAS SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
458099 COAGUCHEK ® XS SYSTEM PROTHROMBIN TIME TEST STRIPS GJS ROCHE DIAGNOSTICS NA 22307421

Patients

Seq Age Sex Outcome Treatment
1 047 YR FISH OIL| 2 "VALVE" REPLACEMENTS| COUMADIN| MULTIVITAMIN