FDA Adverse Event Malfunction Summary report: N

MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT

MDR report key: 3981782 · Received August 5, 2014

Report

Report Number
2955842-2014-04728
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 23, 2014
Report Date
July 23, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THE INSTRUMENT'S GRIP CABLE TO BE BROKEN AT THE DISTAL IDLER. THE IDLER PULLEY SPUN FREELY AND DID NOT EXHIBIT DAMAGE. THE CABLE SEGMENT WAS OBSERVED TO BE STICKING OUT AT THE WRIST OF THE INSTRUMENT. THE OTHER CABLES AT THE WRIST OF THE INSTRUMENT WERE UNDAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE BROKEN CABLE IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT IN CENTRAL PROCESSING, EXPOSED CABLES WERE NOTICED ON THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT. THERE WAS NO REPORT OF FRAGMENTS FALLING INTO A PATIENT. THERE WAS NO PATIENT HARM, ADVERSE OUTCOME OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
458656 MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10130703 425

Patients

Seq Age Sex Outcome Treatment
1