FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3981711 · Received August 5, 2014

Report

Report Number
2032227-2014-06195
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 8, 2014
Report Date
July 8, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE RECEIVED WITH A MOTOR ERROR ALARM DURING REWIND DUE TO CORRODED MOTOR HOME SWITCH .NO A33 ALARM. SCRATCHED LCD WINDOW, CRACKED BATTERY TUBE THREADS AND A CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED DURING THE MANUAL PRIME. CUSTOMER'S CURRENT BLOOD GLUCOSE READING WAS REPORTED TO BE HIGH. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
456756 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522LWWB

Patients

Seq Age Sex Outcome Treatment
1