FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3981654 · Received August 5, 2014

Report

Report Number
2032227-2014-06032
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 5, 2014
Report Date
July 5, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP HAD A NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. INSULIN PUMP RECEIVED WITH CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, CRACKED DISPLAY WINDOW AND CRACKED CASE NEAR DISPLAY WINDOW CORNERS NOTED DURING VISUAL INSPECTION.

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT IS REPORTED THAT THE CUSTOMER'S INSULIN PUMP ALARMED A BUTTON ERROR. CUSTOMER STATED THE INSULIN PUMP WAS SUBMERGED IN WATER FOR TWENTY MINUTES. THE BLOOD GLUCOSE READING WAS 175MG/DL. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
456286 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAL

Patients

Seq Age Sex Outcome Treatment
1 22 YR