FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3981546 · Received August 5, 2014

Report

Report Number
2531779-2014-22465
Event Type
Malfunction
Date Received
August 5, 2014
Report Date
July 26, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/11/2014 WITH THE FOLLOWING: A REVIEW OF THE PUMP¿S BLACK BOX HISTORY SHOWED THAT CS 064 AND CS 078 CALL SERVICE ALARMS WERE RECORDED. DURING TESTING, THE PUMP PERFORMED THE REWIND, LOAD, AND PRIME STEPS WITH NO ALARMS OCCURRING. THE PUMP WAS EXERCISED FOR 24 HOURS ON A 1 UNIT PER HOUR BASAL RATE WITH NO ALARMS OCCURRING. THE CALL SERVICE ALARM ISSUE WAS OBSERVED IN THE PUMP HISTORY BUT WAS NOT ABLE TO BE DUPLICATED DURING INVESTIGATION.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A CALL SERVICE ALARM (CALL SERVICE ALARM ISSUE) ISSUE. THERE WAS NO FURTHER INFORMATION AVAILABLE REGARDING THE COMPLAINT AVAILABLE AT THIS TIME; IF FURTHER INFORMATION IS PROVIDED A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
456810 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 7 YR