4MM FLUTED BALL
Report
- Report Number
- 1045834-2013-16161
- Event Type
- Malfunction
- Date Received
- December 18, 2013
- Date of Event
- August 10, 2012
- Report Date
- August 20, 2012
- Manufacturer
- DEPUY SYNTHES POWER TOOLS
- Product Code
- HBC
- PMA / PMN Number
- K011444
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE DEVICE HAS BEEN RECEIVED BY DEPUY SYNTHES POWER TOOLS. THE DEVICE WAS EVALUATED AND THE CUSTOMER'S COMPLAINT OF PACKAGING DEFECTS COULD NOT BE CONFIRMED. VISUAL INSPECTION ACCEPTANCE CRITERIA ARE "NO LOOSE FOREIGN MATERIAL (LFM) WHEN INSPECTED AT 1X MAGNIFICATION." NO LFM COULD BE FOUND ON DEVICE WHEN INSPECTED AT 10X MAGNIFICATION. ALSO, THE PACKAGE IS IN GOOD CONDITION WITH NO DEFECTS OF THE PEELABLE OR TERMINAL SEAL. WHEN THE INVESTIGATION HAS BEEN COMPLETED OR ADDITIONAL INFO BECOMES AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SENT. REF: (B)(4).
REPORT 1 OF 2. REPORT RECEIVED FROM (B)(6) STATING THAT THE DEVICE HAD DEBRIS IN STERILE PACKAGE. THE DEVICE WAS NOT BEING USED DURING SURGERY. NO INJURY OR MEDICAL INTERVENTION OCCURRED. NO ADDITIONAL INFO PROVIDED. THE DATE OF THE EVENT WAS UNK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 663314 | 4MM FLUTED BALL | HBC | DEPUY SYNTHES POWER TOOLS | F033059521 |