FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3981386 · Received August 5, 2014

Report

Report Number
2032227-2014-06106
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 5, 2014
Report Date
July 5, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP POWERED UP PROPERLY AFTER BATTERY INSTALLATION. NO SHUTTING OFF AFTER BATTERY INSTALLATION ANOMALY WAS NOTED. THE INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS WERE NOTED. NO TRACES OF MOISTURE WERE NOTED AT ELECTRONIC OR MOTOR ASSEMBLIES PER VISUAL INSPECTION. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED CASE AT DISPLAY WINDOW CORNERS. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHES ON THE DISPLAY WINDOW AND A BROKEN RESERVOIR TUBE LIP.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

CUSTOMER CALL REPORTING HER INSULIN PUMP KEEPS ALARMING BUTTON ERROR. THIS HAS NEVER HAPPENED BEFORE. IT WILL TAKE HER FOUR TO FIVE HOURS BEFORE THE DEVICE CAN BE RESET. CUSTOMER'S BLOOD GLUCOSE IS 134 MG/DL. THERE WERE NO SIGNIFICANT EVENTS LEADING TO THE ALARM. CUSTOMER WAS ADVISED THE DEVICE NEEDS TO BE REPLACED, AND TO DISCONTINUE USE AND REVERT TO A BACK UP PLAN. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
458321 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723LNAL

Patients

Seq Age Sex Outcome Treatment
1 36 YR