FDA Adverse Event Injury Summary report: N

SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

MDR report key: 3981373 · Received August 5, 2014

Report

Report Number
1416980-2014-25374
Event Type
Injury
Date Received
August 5, 2014
Date of Event
June 30, 2014
Report Date
July 11, 2014
Manufacturer
BAXTER HEALTHCARE
Product Code
KDJ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ID, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CAUSE OF THIS PERITONITIS WAS USE ERROR DESCRIBED AS BREAK IN ASEPTIC TECHNIQUE. PER BAXTER LABELING, USERS ARE INSTRUCTED TO USE ASEPTIC TECHNIQUE WHEN PERFORMING PERITONEAL DIALYSIS THERAPY. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT HAD A BREAK IN ASEPTIC TECHNIQUE DURING PERITONEAL DIALYSIS (PD) THERAPY RESULTING IN BACTERIAL PERITONITIS. THE PERITONITIS WAS MANIFESTED BY ABDOMINAL PAIN, CLOUDY EFFLUENT AND WEAKNESS. IT WAS NOT REPORTED IF THE PATIENT WAS HOSPITALIZED FOR THE EVENT. THE PATIENT WAS TREATED WITH CEFAZOLIN (1GRAM, DAILY, FOR 13 DAYS, ROUTE NOT REPORTED) FOR PERITONITIS. THE PATIENT WAS RETRAINED ON ASEPTIC TECHNIQUE AND WAS RECOVERING FROM THE PERITONITIS. DIANEAL THERAPY WAS ONGOING. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
458026 SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1 76 YR Required Intervention DIANEAL PD4 AMBUFLEX