FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3981070
·
Received July 24, 2014
Report
- Report Number
- 1720753-2014-06334
- Event Type
- Malfunction
- Date Received
- July 24, 2014
- Date of Event
- July 10, 2014
- Report Date
- July 24, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ID, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE BATTERY CONNECTIONS WERE EVALUATED AND RESEATED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED A PRECHARGE VOLTAGE ERROR MESSAGE. THIS ERROR WILL PREVEN THE SYSTEM FROM BOOTING, RESULTING IN A LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 434839 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |