FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 3981048
·
Received July 24, 2014
Report
- Report Number
- 1720753-2014-06331
- Event Type
- Malfunction
- Date Received
- July 24, 2014
- Date of Event
- July 3, 2014
- Report Date
- July 24, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE INTERCONNECT CABLE WAS EVALUATED AND RESEATED. THE VOLTAGE ON PS1 WAS EVALUATED AND READJUSTED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED LIVE FLUORO IMAGES GO BLACK. THIS RESULTED IN AN INTERMITTENT, RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THIS IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.