FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3980992 · Received July 24, 2014

Report

Report Number
8020893-2014-01804
Event Type
Malfunction
Date Received
July 24, 2014
Date of Event
January 1, 2014
Report Date
July 7, 2014
Manufacturer
COVIDIEN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) INSPECTED THE DEVICE AND VERIFIED THE MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB), BACKLIGHT INVERTER PRINTED CIRCUIT BOARD (PCBS) AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE CSE PERFORMED EXTENDED SELF-TESTING ON THE DEVICE AND ALL TESTS PASSED. (B)(4).

Description of Event or Problem · 1

COVIDIEN RECEIVED INFORMATION STATING THAT THE LOWER DISPLAY ON AN 840 VENTILATOR IS FLICKERING. THE VENTILATOR WAS NOT IN USE ON A PATIENT AT THE TIME THE MALFUNCTION OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
435182 840 VENTILATOR CONTINUOUS VENTILATOR CBK COVIDIEN 840

Patients

Seq Age Sex Outcome Treatment
1