HOMECHOICE PRO
Report
- Report Number
- 1416980-2014-25353
- Event Type
- Malfunction
- Date Received
- August 5, 2014
- Date of Event
- July 16, 2014
- Report Date
- July 16, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE HOMECHOICE (HC) DEVICE WAS RETURNED AND EVALUATED BY THE PRODUCT ANALYSIS LAB (PAL). THE EXTERNAL/INTERNAL INSPECTION WAS PERFORMED WITH NO ISSUES NOTED. TEMPERATURE WAS WITHIN SPECIFICATIONS. THE DEVICE PASSED ELECTRICAL TESTING. A VOLUMETRIC ACCURACY TEST WAS PERFORMED AND THE DEVICE FAILED. TEST ARTICLE, PISTON FOAM, WAS INSTALLED AND THE DEVICE PASSED THE VOLUMETRIC ACCURACY TEST. THE EVALUATION REVEALED THE CAUSE OF THE FAILURE TO BE DETERIORATED PISTON FOAM. THE PISTON FOAM WAS TO BE SCRAPPED AND THE DEVICE WAS SENT FOR SERVICING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE (HC) DEVICE, A BAXTER TECHNICIAN DETERMINED THE HC MACHINE FAILED ACCURACY CONFIRMATION TESTING. DEVICE FAILED DURING EVALUATION, NO PATIENT INVOLVED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 457290 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |