FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3980981 · Received August 5, 2014

Report

Report Number
3004209178-2014-89989
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
May 25, 2014
Report Date
May 25, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP WAS RECEIVED WITH INTERMITTENT BUTTON RESPONSE AND BUTTON ERROR ALARM DUE TO CORRODED KEYPAD TRACES. NO MOISTURE DAMAGE FOUND ON ELECTRONIC ASSEMBLY OR MOTOR. DEVICE HAD SCRATCHED LCD WINDOW, BROKEN BELT CLIP SLOT AT BATTERY TUBE THREADS AREA AND BROKEN RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP RECEIVED A BUTTON ERROR ALARM. CUSTOMER STATED HE IS A LIFE GUARD AND DEVICE WAS EXPOSED TO MOISTURE. BLOOD GLUCOSE LEVEL 123 MG/DL. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
457255 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 17 YR