FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3980976 · Received August 5, 2014

Report

Report Number
3004209178-2014-89987
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
May 24, 2014
Report Date
July 3, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARM NOTED. DEVICE RECEIVED WITH SCRATCHED LCD WINDOW, CRACKED CASE DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT THE INSULIN PUMP RECEIVED A BUTTON ERROR ALARM AND IS UNABLE TO CLEAR IT. THE ONLY EVENT THE CUSTOMER RECALLS LEADING TO THE ALARM IS THAT IT IS EXPOSED TO BODY SWEAT BUT NO CONDENSATION NOTED IN THE INSULIN PUMP. BLOOD GLUCOSE LEVEL IS UNKNOWN.. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
457139 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-523NAS

Patients

Seq Age Sex Outcome Treatment
1 43 YR