FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 3980972
·
Received July 24, 2014
Report
- Report Number
- 8020893-2014-01799
- Event Type
- Malfunction
- Date Received
- July 24, 2014
- Date of Event
- July 2, 2014
- Report Date
- July 3, 2014
- Manufacturer
- NEWPORT MEDICAL INSTRUMENT
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB) AND BACKLIGHT INVERTER PRINTED CIRCUIT BOARDS (PCB). THE CSE UPGRADED THE SOFTWARE TO THE CURRENT REVISION AND PERFORMED THE EXTENDED SELF-TEST (EST), SHORT SELF-TEST (SST) AND THE PERFORMANCE VERIFICATION TEST (PVT); THE DEVICE OPERATED WITHIN THE MANUFACTURING SPECIFICATIONS. (B)(4).
Description of Event or Problem · 1
COVIDIEN RECEIVED INFORMATION STATING THAT AN 840 VENTILATOR HAD IRREGULAR COLORS AND FLICKERING DISPLAY WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 434995 | 840 VENTILATOR | CONTINUOUS VENTILATOR | CBK | NEWPORT MEDICAL INSTRUMENT | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |