FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3980972 · Received July 24, 2014

Report

Report Number
8020893-2014-01799
Event Type
Malfunction
Date Received
July 24, 2014
Date of Event
July 2, 2014
Report Date
July 3, 2014
Manufacturer
NEWPORT MEDICAL INSTRUMENT
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PRINTED CIRCUIT BOARD (PCB) AND BACKLIGHT INVERTER PRINTED CIRCUIT BOARDS (PCB). THE CSE UPGRADED THE SOFTWARE TO THE CURRENT REVISION AND PERFORMED THE EXTENDED SELF-TEST (EST), SHORT SELF-TEST (SST) AND THE PERFORMANCE VERIFICATION TEST (PVT); THE DEVICE OPERATED WITHIN THE MANUFACTURING SPECIFICATIONS. (B)(4).

Description of Event or Problem · 1

COVIDIEN RECEIVED INFORMATION STATING THAT AN 840 VENTILATOR HAD IRREGULAR COLORS AND FLICKERING DISPLAY WHILE IN USE ON A PATIENT. THE PATIENT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
434995 840 VENTILATOR CONTINUOUS VENTILATOR CBK NEWPORT MEDICAL INSTRUMENT 840

Patients

Seq Age Sex Outcome Treatment
1