PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-89857
- Event Type
- Malfunction
- Date Received
- August 5, 2014
- Date of Event
- July 4, 2014
- Report Date
- July 4, 2014
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- OTHER
Narratives
THE INSULIN PUMP PASSED THE DISPLACEMENT, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY ALARM TESTS. THE MOTOR WAS TESTED OUTSIDE OF THE DEVICE AND PASSED. THE INSULIN PUMP WAS RECEIVED WITH A CRACKED RESERVOIR TUBE LIP, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON THE DISPLAY WINDOW AND A SCRATCHED RESERVOIR TUBE WINDOW.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.
IT WAS REPORTED THE INSULIN PUMP ALARMED THREE MOTOR ERRORS. THE CUSTOMER'S BLOOD GLUCOSE LEVEL IS UNKNOWN. CUSTOMER REPORTED THE ALARMS DID NOT OCCUR BECAUSE OF AN MRI SCAN OR SENSOR USE. THE DEVICE IS ABLE TO REWIND. NO FURTHER INFORMATION REPORTED.