FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3980816
·
Received August 5, 2014
Report
- Report Number
- 3004209178-2014-89853
- Event Type
- Malfunction
- Date Received
- August 5, 2014
- Date of Event
- July 4, 2014
- Report Date
- July 4, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP PASSED FUNCTIONAL TESTING INCLUDING REWIND, BASIC OCCLUSION, OCCLUSION, PRIME A/33, EXCESSIVE NO DELIVERY ALARM AND DISPLACEMENT TESTS. NO MOTOR ERROR ALARM NOTED, NO REWIND ANOMALY NOTED. THE INSULIN PUMP WAS RECEIVED WITH MINOR SCRATCHED DISPLAY WINDOW, CRACKED CASE NEAR DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBES THREADS AND CRACKED RESERVOIR TUBE LIP.
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Description of Event or Problem · 1
CUSTOMER REPORTED MOTOR ERROR ALARM. NOTHING FURTHER REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 457635 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-554WWB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |