FDA Adverse Event Malfunction Summary report: N

5CM HEAVY DUTY SHORT ATTACHMENT

MDR report key: 3980731 · Received August 5, 2014

Report

Report Number
1045834-2014-12556
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 23, 2014
Report Date
July 23, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
HBC
PMA / PMN Number
PK011444
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. RELIABILITY ENGINEERING EVALUATED THE DEVICE. AN ASSESSMENT WAS PERFORMED WHICH DETERMINED THAT THE UNIT HAD COME APART. IT WAS OBSERVED THAT THE NOSE TUNE HAD DETACHED FROM THE NOSE CONE, WHICH WAS MOST LIKELY DUE TO THE LOCK TITE BREAKING DOWN OVER TIME. THEREFORE, THE REPORTED CONDITION WAS CONFIRMED. THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE THE DEVICE WORN FROM NORMAL USE AND SERVICING OVER TIME. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING CLEANING IN THE DECONTAMINATION AREA, IT WAS DISCOVERED THAT THE ATTACHMENT DEVICE CAME APART IN TWO PIECES. THE EVENT WAS NOT RELATED TO SURGERY. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
458601 5CM HEAVY DUTY SHORT ATTACHMENT MOTOR, DRILL, ELECTRIC - ATTACHMENT HBC DEPUY SYNTHES POWER TOOLS NA

Patients

Seq Age Sex Outcome Treatment
1