FDA Adverse Event Malfunction Summary report: N

GUARDIAN REAL-TIME MONITIOR

MDR report key: 3980715 · Received August 5, 2014

Report

Report Number
2032227-2014-05948
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 14, 2014
Report Date
July 14, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS RECEIVED WITH BUTTON ERROR ALARM. THERE WAS NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. THE INSULIN PUMP WAS RECEIVED WITH SCRATCHED LCD WINDOW AND MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER'S BLOOD GLUCOSE WAS 136 MG/DL. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
457917 GUARDIAN REAL-TIME MONITIOR CGM MDS MEDTRONIC MINIMED CSS7100NA

Patients

Seq Age Sex Outcome Treatment
1 54 YR