FDA Adverse Event Malfunction Summary report: N

JAGWIRE?

MDR report key: 3980518 · Received August 5, 2014

Report

Report Number
3005099803-2014-02671
Event Type
Malfunction
Date Received
August 5, 2014
Date of Event
July 9, 2014
Report Date
July 10, 2014
Manufacturer
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
Product Code
EZB
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A VISUAL EXAMINATION OF THE GUIDEWIRE REVEALED THAT THE DISTAL TIP WAS DETACHED EXPOSING THE TIP OF THE METAL COREWIRE. PRESENCE OF ADHESIVE REMNANTS WERE FOUND INDICATING THAT THE PEBAX WAS PROPERLY ATTACHED TO THE COREWIRE. NO EVIDENCE OF COREWIRE FRACTURED. THE COMPLAINT IS CONSISTENT WITH THE RETURN THAT THE DISTAL TIP WAS DETACHED EXPOSING THE TIP OF THE METAL COREWIRE. BASED ON ALL GATHERED INFORMATION, THE MOST PROBABLE ROOT CAUSE IS "HANDLING DAMAGE." SINCE THE MANUFACTURING OF THIS COMPLAINT DEVICE, A CORRECTIVE ACTION HAS BEEN INITIATED TO ADDRESS THIS ISSUE. A DHR (DEVICE HISTORY RECORD) REVIEW WAS PERFORMED AND NO DEVIATION WAS FOUND.

Additional Manufacturer Narrative · 1

REPORTED EVENT OF GUIDEWIRE TIP DETACHED EXPOSING THE COREWIRE TIP. THE DEVICE HAS BEEN RECEIVED FOR ANALYSIS. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED.

Description of Event or Problem · 1

NOTE: THIS REPORT PERTAINS TO ONE OF TWELVE DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT #3005099803-2014-02671, 3005099803-2014-02672, 3005099803-2014-02673, 3005099803-2014-02674, 3005099803-2014-02675, 3005099803-2014-02676, 3005099803-2014-02677, 3005099803-2014-02678, 3005099803-2014-02679, 3005099803-2014-02680, 3005099803-2014-02681 AND 3005099803-2014-02682 FOR THE OTHER ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE GUIDEWIRE WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE HYDROPHILIC TIP OF THE JAGWIRE GUIDEWIRE DETACHED INSIDE THE PATIENT EXPOSING THE TIP OF THE METAL COREWIRE. A SECOND JAGWIRE GUIDEWIRE WAS USED, HOWEVER, THE HYDROPHILIC TIP ALSO DETACHED INSIDE THE PATIENT EXPOSING THE TIP OF THE METAL COREWIRE. THE DETACHED TIPS WERE SUCCESSFULLY RETRIEVED USING A BALLOON. TEN MORE JAGWIRE GUIDEWIRES WERE OPENED AND DURING PREPARATION, THE HYDROPHILIC TIP DETACHED EXPOSING THE TIP OF THE METAL COREWIRE. THE PROCEDURE WAS COMPLETED USING THE THIRTEENTH JAGWIRE GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

Description of Event or Problem · 1

NOTE: THIS REPORT PERTAINS TO ONE OF TWELVE DEVICES USED DURING THE SAME PROCEDURE. REFER TO MANUFACTURER REPORT #3005099803-2014-02671, 3005099803-2014-02672, 3005099803-2014-02673, 3005099803-2014-02674, 3005099803-2014-02675, 3005099803-2014-02676, 3005099803-2014-02677, 3005099803-2014-02678, 3005099803-2014-02679, 3005099803-2014-02680, 3005099803-2014-02681 AND 3005099803-2014-02682 FOR THE OTHER ASSOCIATED DEVICE INFORMATION. IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A JAGWIRE GUIDEWIRE WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURE PERFORMED ON (B)(6) 2014. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE HYDROPHILIC TIP OF THE JAGWIRE GUIDEWIRE DETACHED INSIDE THE PATIENT EXPOSING THE TIP OF THE METAL COREWIRE. A SECOND JAGWIRE GUIDEWIRE WAS USED, HOWEVER, THE HYDROPHILIC TIP ALSO DETACHED INSIDE THE PATIENT EXPOSING THE TIP OF THE METAL COREWIRE. THE DETACHED TIPS WERE SUCCESSFULLY RETRIEVED USING A BALLOON. TEN MORE JAGWIRE GUIDEWIRES WERE OPENED AND DURING PREPARATION, THE HYDROPHILIC TIP DETACHED EXPOSING THE TIP OF THE METAL COREWIRE. THE PROCEDURE WAS COMPLETED USING THE THIRTEENTH JAGWIRE GUIDEWIRE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
456976 JAGWIRE? STYLET FOR CATHETER, GASTRO-UROLOGY EZB BOSTON SCIENTIFIC - COSTA RICA (COYOL) M00556581 16080735

Patients

Seq Age Sex Outcome Treatment
1 60 YR