ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2014-05469
- Event Type
- Malfunction
- Date Received
- August 5, 2014
- Date of Event
- July 21, 2014
- Report Date
- July 29, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED WITH TISSUE PAD DETACHED BUT WITH EVIDENCE OF THE TISSUE PAD MATERIAL IN THE GROOVE SECTION OF THE CLAMP ARM. THE DEVICE WAS CONNECTED TO A TEST HANDPIECE AND GENERATOR AND THE DEVICE DID ACTIVATE DURING FUNCTIONAL TESTING. BASED ON THE CONDITION OF THE TISSUE PAD, A PROBABLE CAUSE FOR THIS DAMAGE IS THAT THE CLAMP OF THE DEVICE MAY HAVE BEEN CLOSED AND THE INSTRUMENT ACTIVATED WITHOUT TISSUE PRESENT. CARE SHOULD BE TAKEN NOT TO APPLY PRESSURE BETWEEN THE INSTRUMENT BLADE AND TISSUE PAD WITHOUT HAVING TISSUE BETWEEN THEM. KEEP THE CLAMP ARM OPEN WHEN BACK-CUTTING OR WHILE THE BLADE IS ACTIVE WITHOUT TISSUE BETWEEN THE BLADE AND TISSUE PAD TO AVOID DAMAGE TO THE TISSUE PAD. THE RESULTING DAMAGE CONTRIBUTES TO THE REMOVAL OF THE PAD FROM THE CLAMP ARM. CLEANING OF THE PAD, NOT IN ACCORDANCE WITH THE IFU CAN ALSO RESULT IN REMOVAL OF THE PAD DURING USE.
(B)(4). NO DEVICE RECEIVED FOR ANALYSIS AT TIME OF SUBMISSION OF 3500A. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC SLEEVE GASTRECTOMY THE TISSUE PAD FROM THE DEVICE DISLODGED INTO THE PATIENT. THE TISSUE PAD WAS RETRIEVED FROM THE PATIENT. IT WAS NOT REPORTED HOW THE CASE WAS COMPLETED BUT THERE WAS NO CONSEQUENCE TO THE PATIENT. .
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 456684 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | K9193W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |