ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-22408
- Event Type
- Malfunction
- Date Received
- August 5, 2014
- Report Date
- July 22, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE PUMP HISTORY WAS REVIEWED AND NO EVIDENCE OF A SHORT BATTERY LIFE ISSUE WAS OBSERVED. THE BATTERY VOLTAGES IN THE BLACK BOX WERE WITHIN NORMAL SPECIFICATIONS. TYPICAL USAGE ALARMS WERE OBSERVED IN THE PUMP'S ALARM HISTORY. AN EXTERNAL POWER SUPPLY WAS USED TO TEST THE IDLE, SLEEP, REWIND AND LOAD CURRENTS, AND ALL WERE FOUND TO BE WITHIN SPECIFICATION. NO INTERMITTENT CONNECTIONS OR MOISTURE DAMAGE WAS OBSERVED INSIDE THE PUMP WHEN THE COVER WAS REMOVED FOR INVESTIGATION. THE ALLEGED COMPLAINT COULD NOT BE DUPLICATED. THE BATTERY CAP WAS NOT RETURNED, SO A TEST CAP WAS USED TO COMPLETE ALL TESTING.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE PUMP WAS EXPERIENCING SHORTER THAN EXPECTED BATTERY LIFE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 457123 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR |