FDA Adverse Event
Malfunction
Summary report: N
TRILOGY 100
MDR report key: 3980196
·
Received July 2, 2014
Report
- Report Number
- 2518422-2014-01107
- Event Type
- Malfunction
- Date Received
- July 2, 2014
- Date of Event
- June 12, 2014
- Report Date
- June 12, 2014
- Manufacturer
- RESPIRONICS INC.,
- Product Code
- CBK
- PMA / PMN Number
- K083526
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND THE CUSTOMER'S COMPLAINT WAS CONFIRMED, THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUE. DAMAGE CONSISTENT WITH THE DEVICE BEING DROPPED WAS OBSERVED. CONCLUSIONS: DEVICE HAS BEEN EVALUATED BUT NOT REPAIRED, PENDING CUSTOMER APPROVAL OF THE ESTIMATE.
Description of Event or Problem · 1
THE MANUFACTURER RECEIVED INFORMATION ALLEGING A VENTILATOR FAILED A STEP DURING TESTING. THE DEVICE WAS NOT IN PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 387966 | TRILOGY 100 | VENTILATOR, CONTINUOUS, FACILITY USE | CBK | RESPIRONICS INC., | 1054096 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |