FDA Adverse Event
Malfunction
Summary report: N
9600
MDR report key: 3980185
·
Received June 5, 2014
Report
- Report Number
- 1720753-2014-04755
- Event Type
- Malfunction
- Date Received
- June 5, 2014
- Date of Event
- May 14, 2014
- Report Date
- June 5, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE COLLIMATOR WAS EVALUATED AND REPAIRED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM EXHIBITED 'IRIS POTENTIOMETER ERRORS' UPON SYSTEM START UP. THIS ERROR WILL PREVENT THE SYSTEM FROM BOOTING TO A USABLE STATE. NO PATIENT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 330416 | 9600 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |