FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3980165
·
Received July 2, 2014
Report
- Report Number
- 1720753-2014-05629
- Event Type
- Malfunction
- Date Received
- July 2, 2014
- Date of Event
- June 16, 2014
- Report Date
- July 2, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE FAILURE COULD NOT BE DUPLICATED. THE COIN BATTERIES ON RTOS, GPOS, AND GIB WERE REPLACED. THE WORKSTATION AND MAINFRAME PCBS WERE RESEATED. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYS SHUT DOWN WITHOUT COMMAND. THIS RESULTED IN A LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 388798 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |