FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3980082 · Received July 2, 2014

Report

Report Number
1720753-2014-05645
Event Type
Malfunction
Date Received
July 2, 2014
Date of Event
June 17, 2014
Report Date
July 2, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE INTERCONNECT CABLE AND HARD DISK DRIVE WERE EVALUATED AND REPLACED. THE SYS SOFTWARE AND CALIBRATIONS WERE ALSO RELOADED DURING THE SERVICE EVENT. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYS INTERMITTENTLY FAILED TO PROPERLY SAVE AND RECALL PT IMAGE DATA. NO PT SERIOUS INJURY OR DEATH WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
388778 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1