FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3980044 · Received June 5, 2014

Report

Report Number
1720753-2014-04729
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
May 16, 2014
Report Date
June 5, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE POWER CAP MODULE CHARGER BOARD AND BATTERY PACK ASSY WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A PRECHARGE VOLTAGE ERROR MESSAGE. THIS ERROR WILL PREVENT THE SYSTEM FROM BOOTING, RESULTING IN A LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
330458 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1