FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3980012 · Received June 5, 2014

Report

Report Number
1720753-2014-04719
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
May 15, 2014
Report Date
June 5, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE PS2, FAST SWITCH AND CABLE WERE EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THERE WAS A FAST STOP ACTIVATED ERROR MESSAGE. THIS ERROR WILL CAUSE THE SYSTEM TO LOCK UP AND HAVE TO BE REBOOTED. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
330326 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1