FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3980007 · Received June 5, 2014

Report

Report Number
1720753-2014-04716
Event Type
Malfunction
Date Received
June 5, 2014
Date of Event
May 15, 2014
Report Date
June 5, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE GIB (GENERATOR INTERFACE BOARD) WAS EVALUATED AND REPLACED. THE SYSTEM SOFTWARE WAS RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER AND THE FSE REPORTED MULTIPLE ERRORS THAT CAUSED THE SYSTEM TO INTERMITTENTLY SHUT DOWN. THIS RESULTED IN AN INTERMITTENT, RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
330365 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1