FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 3980007
·
Received June 5, 2014
Report
- Report Number
- 1720753-2014-04716
- Event Type
- Malfunction
- Date Received
- June 5, 2014
- Date of Event
- May 15, 2014
- Report Date
- June 5, 2014
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE GIB (GENERATOR INTERFACE BOARD) WAS EVALUATED AND REPLACED. THE SYSTEM SOFTWARE WAS RELOADED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND RETURNED TO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER AND THE FSE REPORTED MULTIPLE ERRORS THAT CAUSED THE SYSTEM TO INTERMITTENTLY SHUT DOWN. THIS RESULTED IN AN INTERMITTENT, RECOVERABLE LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 330365 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |