FDA Adverse Event Malfunction Summary report: N

CS300

MDR report key: 3979735 · Received July 2, 2014

Report

Report Number
2249723-2014-00975
Event Type
Malfunction
Date Received
July 2, 2014
Date of Event
January 24, 2011
Report Date
January 24, 2011
Manufacturer
DATASCOPE CORP
Product Code
DSP
PMA / PMN Number
K063525
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REP REPLACED THE DISPLAY CONTROLLER BOARD (PART NUMBER 0670-00-0640E), DISPLAY CABLE CONNECTOR (PART 0012-00-1058), IV POLE SHAFT COLLAR (PART NUMBER 0105-00-0114), BUSHING POLE MOUNT (PART NUMBER 0358-00-0063) WERE REPLACED. EXPIRED SAFETY DISC ASSEMBLY (PART NUMBER 0997-00-0985-01), BRACKET LOCKING CART (PART NUMBER 0406-00-0860), LATCH CONSOLE RELEASE CART (PART NUMBER 0367-00-0091) AND SCREW SHOULDER (PART NUMBER 0212-00-0096) WERE REPLACED. THE COMPANY REP PERFORMED PNEUMATIC PERFORMANCE SERVICE DIAGNOSTICS, FUNCTIONAL TESTS, ELECTRICAL SAFETY CHECKS, CALIBRATION AND FULL PREVENTIVE MAINTENANCE. THE IABP MET FACTORY SPECS AND WAS RETURNED TO THE CUSTOMER. (B)(4).

Description of Event or Problem · 1

DURING A PREVENTIVE MAINTENANCE, THE COMPANY REP OBSERVED FLUID SPILL ON THE DISPLAY CONTROLLER BOARD AND ON THE DISPLAY CABLE CONNECTOR. IV POLE SHAFT COLLAR WAS FOUND BROKEN AND BUSHING POLE MOUNT WAS MISSING. THE HELIUM TANK GAUGE READING WAS 1852 PSI. NO PT WAS INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
387042 CS300 INTRA-AORTIC BALLOON PUMP DSP DATASCOPE CORP CS300

Patients

Seq Age Sex Outcome Treatment
1