APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR

MDR report key: 3978987 · Received June 27, 2014

Report

Report Number
3003673482-2014-00013
Event Type
Malfunction
Date Received
June 27, 2014
Date of Event
May 2, 2014
Report Date
May 2, 2014
Manufacturer
LIFE TECHNOLOGIES HOLDINGS PTE. LTD,
Product Code
NSU
PMA / PMN Number
K082562
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED / EXPLANTED. THE INSTRUMENT WAS INVESTIGATED ON SITE. PROBLEM SOLVED ON SITE BY REPLACING THE USB ADAPTOR. DEVICE INTENDED USE: THE APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT WITH THE (B)(4) SOFTWARE VERSION 1.4 IS A REAL-TIME NUCLEIC ACID AMPLIFICATION AND DETECTION SYSTEM THAT MEASURES NUCLEIC ACID SIGNALS FROM REVERSE TRANSCRIBED RNA AND CONVERTS THEM TO COMPARATIVE QUANTITATIVE READOUTS USING FLUORESCENT DETECTION OF DUAL-LABELED HYDROLYSIS PROBES. THE 7500 FAST DX REAL-TIME PCR INSTRUMENT IS TO BE USED ONLY BY TECHNOLOGISTS TRAINED IN LAB TECHNIQUES, PROCEDURES, AND ON USE OF THE ANALYZER.

Description of Event or Problem · 1

A CUSTOMER REPORTED THERE WAS CONNECTION ISSUE IN THE INSTRUMENT APPLIED BIOSYSTEMS 7500 FAST DX (CAT # 4407205, SERIAL# (B)(4)) THAT CAUSED THE INSTRUMENT TO STOP IN THE MIDDLE OF AN ASSAY RUN. NO PT INVOLVEMENT REPORTED. (B)(4).