APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR
Report
- Report Number
- 3003673482-2014-00018
- Event Type
- Malfunction
- Date Received
- June 27, 2014
- Date of Event
- March 7, 2013
- Report Date
- March 7, 2013
- Manufacturer
- LIFE TECHNOLOGIES HOLDINGS PTE LTD
- Product Code
- NSU
- PMA / PMN Number
- K082562
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. THE FIELD SVC ENGINEER ATTRIBUTED THE CONTAMINATION ISSUE TO WEAR AND TEAR OF THE HEATED COVER. THE HEATED COVER WAS REPLACED AND THE CONTAMINATION ISSUE WAS RESOLVED. DEVICE INTENDED USE: THE APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT WITH THE SDS SOFTWARE VERSION 1.4 IS A REAL-TIME NUCLEIC ACID AMPLIFICATION AND DETECTION SYSTEM THAT MEASURES NUCLEIC ACID SIGNALS FROM REVERSE TRANSCRIBED RNA AND CONVERTS THEM TO COMPARATIVE QUANTITATIVE READOUTS USING FLUORESCENT DETECTION OF DUAL-LABELED HYDROLYSIS PROBES. THE 7500 FAST DX REAL-TIME PCR INSTRUMENT IS TO BE USED ONLY BY TECHNOLOGISTS TRAINED IN LABORATORY TECHNIQUES, PROCEDURES AND ON USE OF THE ANALYZER.
A (B)(6) FIELD SVC ENGINEER REPORTED THAT CONTAMINATION FROM THE HEATED COVER OF APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR INSTRUMENT (CATALOG NO. 4407205 SERIAL NO. (B)(4)) CAUSED INCORRECT RESULTS TO BE REPORTED FROM A RUN. WE WERE NOT MADE AWARE OF ANY IMPACT TO PT MANAGEMENT. THE COMPANY EVENT TRACKING NO: (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 377149 | APPLIED BIOSYSTEMS 7500 FAST DX REAL-TIME PCR | INSTRUMENTATION FOR CLINICAL MULTIP | NSU | LIFE TECHNOLOGIES HOLDINGS PTE LTD |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |